SenseAir ® S8 – Key technical specification
Item
|
SenseAir® S8
|
Possible functionality of other members in SenseAir® S8 family
|
Target gas
|
CO2
|
Contact SenseAir for other gases availability
|
Measurement range
|
400 to 5000 ppm (Note 1),
Up to 10000ppm extended range (Note 2)
|
|
Accuracy
|
+-30ppm+-3% of reading
(Notes 3 and 4) |
Sensor rated for California standard is planned. Please contact SenseAir for further information
|
Response time
|
2 minutes by 90% for diffusion sampling method model
|
|
Operating temperature
|
0 to 50C
|
Sensors for extended temperature ranges -40 to +85C are under development
|
Operating humidity
|
0 to 95% RH non condensed
|
|
Storage temperature
|
-40 to + 70 C
|
Sensors for extended temperature ranges -40 to +85C are under development
|
Dimensions (body)
|
32.9*19.7*9.9 mm3 for #400010
33.7*19.7*9.9 mm3 for #400011 |
|
Weight
|
TBD
|
|
Power supply
|
5V +-5% unprotected against surges and reverse connection
|
|
Power consumption
|
300 mA peak,
30 mA average |
Low power versions are planned. Contact SenseAir for details and schedule.
|
Sensor life
|
15 years in normal commercial environments with ABC on (Note 3)
|
|
Compliance with
|
CE conformity declaration not required because it’s not a product directed to end user. Our customers have to make this declaration for their entire product.
RoHS directive 2002/95/EG
|
SenseAir is going to perform immunity and emission tests anyway to confirm compliance of the module in immunity and emission
Other tests / approvals are available on request, contact SenseAir
|
Item
|
SenseAir® S8
|
Possible functionality of other members in SenseAir® S8 family
|
Serial communication
|
UART, Modbus protocol (Note 5)
Direction control pin for direct connection to RS485 receiver integrated circuit.
|
FastStream protocol
delivering dedicated measurement data with dedicated time intervals is under development |
Alarm output, Open Collector
|
Normally Conducting after power up,
Non conducting (high level) at > 1000 ppm,
return to conducting (low level) at < 800 ppm
Non Conducting (high level) at error detected
|
Contact SenseAir for other optional functionality of OC and threshold configuration.
|
PWM output, 1 kHz
|
0 to 100% duty cycle for 0 to 2000 ppm
3.3V push-pull CMOS output, unprotected
|
Contact SenseAir for other optional output ranges.
|
Forced calibration
|
Background calibration in fresh air (400 ppm) by forcing bCal input Low for about 4 sec.
Keeping bCal input low for more than 16 sec forces sensor to perform zero (nitrogen) calibration
|
|